Changing the Nebulizer

Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.
Comment: Migrated to Confluence 4.0

Cross flow to ModLichte

1. After unscrewing the metal door, remove to the two metal clips by unscrewing the steel collars.

2. Remove the glass tube that the clips were holding in place.

3. Detach the teflon tube from the plastic tube running outside the machine as well as the plastic cord with the black rubber end (leave the rubber piece attatched to the nozzle), and disconnect the cord labled "M"

4. Remove the nebulizer from the machine and detach the pressure sealed tan cap.

5. Take the glass tube specific to the modlichte and attach one end to the plasma via the first metal clip and attach the other end to the modlichte via the second clip.

6. Attach the teflon tube on the modlichte to the plastic tube where the cross flow's teflon tube was previously removed.

7. Connect the modlichte to the Cord labled "M" and to the plastic tube with the black rubber end.

8. Change the ICal reference settings to 'Modlichte'

Modlichte to Cross Flow

1. After unscrewing the metal door, remove to the two metal clips by unscrewing the steel collars.

2. Remove the glass tube that the clips were holding in place.

3. Detach the teflon tube from the plastic tube running outside the machine as well as the plastic cord with the black rubber end (leave the rubber piece attatched to the nozzle), and disconnect the cord labled "M"

4. Remove the nebulizer from the machine

5. Take the glass tube specific to the cross flow and attach one end to the plasma via the first metal clip and attach the other end to the cross flow via the second clip.

6. Attach the teflon tube on the cross flow to the plastic tube where the modlichte's teflon tube was previously removed.

7. Connect the cross flow to the Cord labled "M" and to the plastic tube with the black rubber end.

8. Change the ICal reference settings to 'Cross flow'